Section 6: SCPI command reference
[:SENSe[1]]:<function>:RANGe:AUTO:ULIMit
When autorange is selected, this command selects the upper limit for measurements of the selected function.
Type
Command and query
Usage
[:SENSe[1]]:<function>:RANGe:AUTO:ULIMit <n>
[:SENSe[1]]:<function>:RANGe:AUTO:ULIMit?
[:SENSe[1]]:<function>:RANGe:AUTO:ULIMit? DEFault
[:SENSe[1]]:<function>:RANGe:AUTO:ULIMit? MINimum
[:SENSe[1]]:<function>:RANGe:AUTO:ULIMit? MAXimum
<function>
<n>
Details
You can use this command when automatic range selection is enabled to put an upper bound on the range that
is used for resistance measurements.
The upper limit must be more than the lower limit.
If the lower limit is equal to the upper limit, automatic range setting is effectively disabled.
Example
:SENSe:RESistance:RANGe:AUTO:ULIMit 20
Also see
[:SENSe[1]]:<function>:RANGe:AUTO
[:SENSe[1]]:<function>:RANGe:AUTO:LLIMit
6-52
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The measurement function to which this setting applies:
Voltage (query only): VOLTage[:DC]
Resistance: RESistance
The upper limit:
Voltage: 0.02 to 200 volts
Resistance: 20 to 2.0e8 ohms
(on page 6-50)
Where saved
Save settings
Measure configuration list
Set the upper limit to 20 ohms.
(on page 6-51)
Default value
Resistance: 2e5