Section 6: SCPI command reference
*RST
SENS:FUNC "CURR"
SENS:CURR:RANG:AUTO ON
SENS:CURR:RSEN OFF
SOUR:FUNC VOLT
SOUR:VOLT:RANG 20
SOUR:VOLT:ILIM 1
SOUR:LIST:VOLT 1, 5, 1, 5, 1, 5
SOUR:SWE:VOLT:LIST 1, 0.2
INIT
*WAI
TRAC:DATA? 1, 6, "defbuffer1", SOUR, READ
Example 2
SOUR:CONF:LIST:CRE "biasLevel"
SOUR:FUNC VOLT
SENS:FUNC "CURR"
SOUR:VOLT:LEV 5
SOUR:CONF:LIST:STORE "biasLevel"
SOUR:SWE:VOLT:LIST 1, .001, 1, 1, "defbuffer2", "biasLevel"
INIT
This example uses a user-defined configuration list.
Create a configuration list named biasLevel. Set the source function to 5 volts and the measure function to
current.
Store the configuration list.
Set up a voltage sweep that uses the configuration list, starting at index point 1 with a delay of 1 ms. The
sweep is to abort if the source limit is exceeded, store data in defbuffer2, and use the configuration list
biasLevel.
Also see
Configuration lists
:INITiate[:IMMediate]
:SOURce[1]:LIST:<function>
Sweep operation
6-86
Model 2450 Interactive SourceMeter® Instrument Reference Manual
(on page 3-33)
(on page 6-124)
(on page 6-78)
(on page 3-51)
This example uses the
:SOURce[1]:LIST:<function>
command to set up the configuration list that
is used by the sweep.
This example will source 1 V, 5 V, 1 V, 5 V,
1 V, 5 V and measure the resulting current
at each voltage point. The time duration of
each voltage point is 200 ms.