Status.questionable.setmap() - Keithley 2450 Reference Manual

Interactive sourcemeter instrument
Table of Contents

Advertisement

Section 8: TSP command reference

status.questionable.setmap()

This attribute maps events to bits in the questionable event registers.
Type
TSP-Link accessible
Function
Yes
Usage
status.questionable.setmap(bitNumber, setEvent)
status.questionable.setmap(bitNumber, setEvent, clearEvent)
bitNumber
setEvent
clearEvent
Details
You can map events to bits in the event registers with this command. This allows you to cause bits in the
condition and event registers to be set or cleared when the specified events occur.
When a mapped event is programmed to set bits, the corresponding bits in both the condition register and event
register are set when the event is detected.
When a mapped event is programmed to clear bits, the bit in the condition register is set to 0 when the event is
detected.
If the event is set to zero (0), the bit is never set.
Also see
status.questionable.getmap()
status.request_enable
This attribute stores the settings of the Service Request (SRQ) Enable Register.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
SRQEnableRegister = status.request_enable
status.request_enable = SRQEnableRegister
SRQEnableRegister
8-160
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Affected by
The bit number that is being mapped to an event
The event that will set the bit
The event that will clear the bit
(on page 8-159)
Affected by
Status reset
The status of the service request (SRQ) enable register; a zero (0)
indicates no bits set (also send 0 to clear all bits); other values indicate
various bit settings (0 to 255)
Where saved
Default value
Where saved
Default value
Not saved