Section 8: TSP command reference

localnode.model

This attribute stores the model number.
Type
TSP-Link accessible
Attribute (R)
Yes
Usage
model = localnode.model
model
Details
When using this command from a remote node, replace localnode with the node reference, for example,
node[5].model.
Example
print(localnode.model)
Also see
localnode.serialno

localnode.password

This attribute stores the instrument password.
Type
TSP-Link accessible
Attribute (W)
No
Usage
localnode.password = "password"
passWord
Details
When the access to the instrument is set to protected or lockout, this is the password that is used to gain access.
The instrument continues to use the old password for all interactions until the command to change it executes.
When changing the password, give the instrument time to execute the command before attempting to use the
new password.
If you forget the password, you can reset the password to the default. On the front panel, press MENU. Under
System, select Manage. Select LAN and Password Reset. You can also reset the password from the rear panel
by inserting a straightened paper clip into hole below LAN RESET.
If you are using this command from a remote node, replace localnode with the node reference, for example,
node[5].password.
Example
localnode.password = "N3wpa55w0rd"
Also see
localnode.access
8-70
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Affected by
Not applicable
The model number of the instrument
(on page 8-72)
Affected by
LAN reset
A string that contains the instrument password (maximum 30 characters)
(on page 8-68)
Where saved
Not applicable
Outputs the model number of the local node. For example:
2450
Where saved
Nonvolatile memory
Changes the password to N3wpa55w0rd.
Default value
Not applicable
Default value
"admin"
2450-901-01 Rev. B/September 2013