Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 2: General operation
Setting the instrument to use the front or rear terminals
If the output is on when you change the settings for the terminals that are used, the output turns off.
Using the front panel:
1. Press the FUNCTION key.
2. Select the source and measurement combination
3. Press the TERMINALS FRONT/REAR switch.
When F is lit, the instrument is using the front-panel terminals. When R is lit, the instrument is using
rear-panel terminals.
Using SCPI commands:
To change to the front-panel terminals for current measurements, send the command:
ROUTe:TERMinals FRONt
To change to the rear-panel terminals for current measurements, send the command:
ROUTe:TERMinals REAR
Using TSP commands:
To change to the front-panel terminals, send the command:
smu.measure.terminals = smu.TERMINALS_FRONT
To change to the rear-panel terminals, send the command:
smu.measure.terminals = smu.TERMINALS_REAR
2450-901-01 Rev. B/September 2013
2-73