Keithley 2450 Reference Manual page 166

Interactive sourcemeter instrument
Table of Contents

Advertisement

Model 2450 Interactive SourceMeter® Instrument Reference Manual
Using the front panel to save or append buffer content to files:
1. Insert a USB drive into the USB port.
2. Press the MENU key.
3. Under Measure, select Data Buffers. The MANAGE MEASURE BUFFERS window is displayed.
4. Select the reading buffer that you want to save. For example, select testData. The Settings for
testData menu is displayed.
5. Select the Save To USB button. A keypad is displayed.
6. Enter the name of the file in which to save the readings.
You only have to enter the name of the file you want to save. It is not necessary to enter the file
extension. All files are saved as .csv files.
7. Select Yes to confirm saving the file. When the MANAGE MEASURE BUFFERS window is
displayed again, the file is saved.
8. Press the HOME key to return to the HOME screen.
2450-901-01 Rev. B/September 2013
Figure 72: MANAGE MEASURE BUFFERS window
Figure 73: MANAGE MEASURE BUFFERS
Section 3: Functions and features
3-21

Advertisement

Table of Contents
loading

Table of Contents