Section 6: SCPI command reference
:SOURce[1]:<function>:DELay
This command sets the source delay.
Type
Command and query
Usage
:SOURce[1]:<function>:DELay <n>
:SOURce[1]:<function>:DELay?
:SOURce[1]:<function>:DELay? DEFault
:SOURce[1]:<function>:DELay? MINimum
:SOURce[1]:<function>:DELay? MAXimum
<function>
<n>
Details
This command sets a delay for the selected source function. This delay is in addition to normal settling times.
After the programmed source is turned on, this delay allows the source level to settle before a measurement is
taken.
If source autodelay is on, if you set a specific delay, it is turned off.
If source autodelay is on, the manual source delay setting is not saved in the source configuration list.
If you send this command without the <function> parameter, it sets the delay for all functions.
Example
SOUR:VOLT:DELay DEF
Also see
:SOURce[1]:<function>:DELay:AUTO
6-66
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Source configuration list
The source function to which this setting applies:
Current: CURRent
•
Voltage: VOLTage
•
The delay in seconds (0 to 4)
(on page 6-67)
Where saved
Save settings
Source configuration list
Set the delay for the voltage source to the
default value.
2450-901-01 Rev. B/September 2013
Default value
Not applicable