Model 2450 Interactive SourceMeter® Instrument Reference Manual
Storing configuration point 4
Use one of the following methods to:
Set the instrument source level to 5.0 V
Store all active source settings to MySourceList as configuration point 4 by appending to the
end of the list
Using SCPI commands:
:SOURce:VOLTage:LEVel 5
:SOURce:CONF:LIST:STORe "MySourceList"
Using TSP commands:
smu.source.level = 5
smu.source.configlist.store("MySourceList")
The following figure shows the active state of the instrument after you change the source level to
5.0 V. Refer to
list of source settings that the instrument stores in a source configuration list.
2450-901-01 Rev. B/September 2013
Instrument settings stored in a source configuration list
Figure 85: Example configuration point 4
Section 3: Functions and features
(on page 3-38) for a complete
3-47