Keithley 2450 Reference Manual page 185

Interactive sourcemeter instrument
Table of Contents

Advertisement

Section 3: Functions and features
The CONFIGURATION LIST menu contains:
A scrollable list of configuration points that are stored in the selected configuration list
A message bar indicating the index associated with the selected configuration point
Buttons to perform the operations, described below
Button
Select
New List
Delete Point
System to List
List to System
View Details
Duplicate configuration points:
If you store a second configuration point that has the same settings as a point that is already on the
configuration list, the "No change" message is displayed.
Creating a configuration list and giving it a name
For example, use the following information to create MyMeasList.
Using the front panel to create the configuration list:
1. Press the MENU key.
2. Under Measure, select Config List. The MEASURE CONFIGURATION LISTS screen is
displayed.
3. Select New List. The keypad is displayed.
4. Enter a name for the configuration list you are creating. For example, MyMeasList.
5. Select the Enter button on the keypad. The MEASURE CONFIGURATION LISTS screen is
displayed.
6. Select HOME to return to the home screen.
Storing configuration point 1
For example, use the following information to store configuration point 1 to MyMeasList.
Using the front panel to configure the instrument:
Configure the instrument with the settings you want to store in the configuration point. For example:
1. Press the QUICKSET key and select the SrcV MeasI function to source voltage and measure
current.
2. Press the MENU key.
3. Under Measure, select Settings. The MEASURE SETTINGS menu is displayed.
4. Select Measure Range, then select 10 nA.
5. Select NPLC, then select 1.00.
3-40
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Use to ...
If you selected Source on the main menu, a menu of the source configuration lists
presently available on the instrument is displayed.
If you selected Measure on the main menu, a menu of the measure configuration lists
presently available on the instrument is displayed.
Create a new configuration list.
Delete the selected configuration point.
Store the active settings into a configuration point on the specified configuration list.
New points are chronologically numbered with the index starting from 1.
Recall a configuration point and restore the stored settings to the active state.
Display the details of the selected configuration point.