Aborting A Sweep - Keithley 2450 Reference Manual

Interactive sourcemeter instrument
Table of Contents

Advertisement

Model 2450 Interactive SourceMeter® Instrument Reference Manual
For example sweeps, see
For detail on the commands and options listed above, see the following command descriptions:
smu.source.sweeplinear()
smu.source.sweeplinearstep()
smu.source.sweeplist()
smu.source.sweeplog()
smu.source.func
smu.source.range
smu.source.configlist.create()
smu.source.configlist.store()
smu.measure.func
smu.measure.range

Aborting a sweep

Sweeps can be stopped for the following reasons:
The limit set by the abort on limit setting was exceeded
The trigger model is aborted
You can stop the sweep while it is in progress. When you stop the sweep, all sweep commands in the
trigger model are terminated.
Using the front panel:
Press the front-panel TRIGGER key for two seconds and select Trigger Model (Abort).
Using SCPI commands:
Send the command:
:ABORt
Using TSP commands:
Send the command:
trigger.model.abort()
2450-901-01 Rev. B/September 2013
Sweep programming examples
(on page 8-145)
(on page 8-147)
(on page 8-149)
(on page 8-151)
(on page 8-134)
(on page 8-143)
(on page 8-129)
(on page 8-132)
(on page 8-103)
(on page 8-117)
Section 3: Functions and features
(on page 3-58).