Section 6: SCPI command reference
:SYSTem:EVENtlog:COUNt?
This command returns the number of events in the event log.
Type
Query only
Usage
:SYSTem:EVENtlog:COUNt?
:SYSTem:EVENtlog:COUNt? <mask>
:SYSTem:EVENtlog:COUNt? <mask>, <mask>
<mask>
Details
You can use the mask parameter to limit the event log items that are counted.
This command does not clear the event log.
If you do not define a mask, all events are counted.
Example
:SYST:EVEN:COUN? ERR
Also see
:SYSTem:CLEar
:SYSTem:EVENtlog:NEXT?
:SYSTem:EVENtlog:SAVE
6-100
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Affected by
Not applicable
Limits the list of event log entries to specific types; set to:
ERRor: Returns the number of errors
•
WARNing: Returns the number of warnings
•
INFormational: Returns the number of informational messages
•
ALL: Returns all events
•
(on page 6-96)
(on page 6-101)
(on page 6-103)
Where saved
Not applicable
Displays the present number of errors in the
instrument event log.
If there are three errors in the event log,
output is:
3
2450-901-01 Rev. B/September 2013
Default value
Not applicable