Model 2450 Interactive SourceMeter® Instrument Reference Manual
:TRIGger:BLENder<n>:CLEar
This command clears the blender event detector and resets the overrun indicator of blender <n>.
Type
Command only
Usage
:TRIGger:BLENder<n>:CLEar
<n>
Details
This command sets the blender event detector to the undetected state and resets the overrun indicator of the
event detector.
Example
:TRIG:BLEN2:CLE
Also see
None
:TRIGger:BLENder<n>:MODE
This command selects whether the blender performs OR operations or AND operations.
Type
Command and query
Usage
:TRIGger:BLENder<n>:MODE <operation>
:TRIGger:BLENder<n>:MODE?
<n>
<operation>
2450-901-01 Rev. B/September 2013
Affected by
Not applicable
The blender number (1 or 2)
Affected by
Recall settings
Instrument reset
Power cycle
Trigger blender clear
The blender number (1 or 2)
The type of operation:
OR
AND
Section 6: SCPI command reference
Where saved
Not applicable
Clears the event detector for blender 2.
Where saved
Save settings
Default value
Not applicable
Default value
AND