Output[1][:State] - Keithley 2450 Reference Manual

Interactive sourcemeter instrument
Table of Contents

Advertisement

Section 6: SCPI command reference

:OUTPut[1][:STATe]

This command enables or disables the source output.
Type
Command and query
Usage
:OUTPut[1][:STATe] <b>
:OUTPut[1][:STATe]?
<b>
Details
When the output is switched on, the instrument sources either voltage or current, as set by
[:SOURce[1]]:FUNCtion[:MODE].
Example
:OUTP ON
Also see
:SOURce[1]:FUNCtion[:MODE]
6-34
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Source configuration list
Turn source off: 0 or OFF
Turn source on: 1 or ON
(on page 6-72)
Where saved
Save settings
Source configuration list
Switch the source output of the instrument to
on.
2450-901-01 Rev. B/September 2013
Default value
0 (OFF)

Advertisement