Section 8: TSP command reference

smu.measure.math.mxb.mfactor

This attribute specifies the scale factor for the y = mx + b math operation.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
value = smu.measure.math.mxb.mfactor
smu.measure.math.mxb.mfactor = value
value
Details
This command sets the scale factor (m) for an mx + b operation for the selected measurement function.
The mx+b math operation lets you manipulate normal display readings (x) mathematically according to the
following calculation:
y = mx + b
Where:
y is the displayed result
m is a user-defined constant for the scale factor
x is the measurement reading (if you are using a relative offset, this is the measurement with relative
offset applied)
b is the user-defined constant for the offset factor
Example
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.math.format = smu.MATH_MXB
smu.measure.math.mxb.mfactor = 0.80
smu.measure.math.mxb.bfactor = 50
smu.measure.math.enable = smu.ON
Also see
Calculations that you can apply to measurements
smu.measure.math.enable
smu.measure.math.mxb.bfactor
8-114
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Affected by
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
The scale factor; the valid range is −4294967295 to +4294967295
(on page 3-6)
(on page 8-111)
(on page 8-113)
Where saved
Default value
Configuration script
1
Measure configuration list
Set the measurement function to voltage.
Set the scale factor for the mx +b operation
to 0.80.
Set the offset factor to 50.
Enable the math function.