Safety Shielding; Grounding; Noise And Chassis Ground - Keithley 2450 Reference Manual

Interactive sourcemeter instrument
Table of Contents

Advertisement

Model 2450 Interactive SourceMeter® Instrument Reference Manual

Safety shielding

Use #18 AWG wire or larger for connections to safety earth ground and chassis.

Grounding

Noise and chassis ground

Using the chassis as a ground point for signal connections to the Model 2450 chassis may result in
different levels of noise, depending on your setup. If the Model 2450 common-mode current is
channeled to the chassis instead of the device, the tie point to the chassis can help quiet
measurements. However, if other equipment is connected to the chassis, you may have more noise
because of other connected equipment.
If you choose to use the chassis as a ground point for signal connections, use the Model 2450
chassis screw as a connection point.
2450-901-01 Rev. B/September 2013
Figure 116: Model 2450 noise and safety shield
Section 4: Source-measure considerations
4-11