Section 8: TSP command reference

smu.measure.limit[Y].autoclear

This command indicates if limit Y should be cleared automatically or not.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
value = smu.measure.limit[Y].autoclear
smu.measure.limit[Y].autoclear = value
value
Y
Details
When this command sets autoclear to on for a measurement function, if a measurement fails limit, but the next
measurement passes limit, the failed limit condition is cleared. Therefore, if you are making a series of
measurements, the instrument uses last measurement limit for the fail indication for the limit.
If you want to know if any of a series of measurements failed the limit, set the auto clear setting to off. When this
set to off, a failed indication is not cleared automatically. It remains set until it is cleared with the clear command.
The auto clear setting affects both the high and low limits of Y.
Example
smu.measure.func = smu.FUNC_DC_CURRENT
smu.measure.limit[1].autoclear = smu.ON
Also see
smu.measure.limit[Y].clear()
8-104
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Affected by
Where saved
Restore configuration
Configuration script
Instrument reset
Measure configuration list
Power cycle
Measure configuration list
Function change
The auto clear setting:
Enable: smu.ON or 1
Disable: smu.OFF or 0
Limit number: 1 or 2
(on page 8-105)
Default value
smu.ON
Turns on autoclear for limit 1 when
measuring DC current.
2450-901-01 Rev. B/September 2013