Increasing The Speed Of Sweeps - Keithley 2450 Reference Manual

Interactive sourcemeter instrument
Table of Contents

Advertisement

Model 2450 Interactive SourceMeter® Instrument Reference Manual
Using TSP commands
reset()
smu.source.configlist.create("CurrListSweep")
smu.source.func = smu.FUNC_DC_CURRENT
smu.source.range = 100e-3
smu.source.level = 1e-3
smu.source.configlist.store("CurrListSweep")
smu.source.level = 10e-3
smu.source.configlist.store("CurrListSweep")
smu.source.level = 5e-3
smu.source.configlist.store("CurrListSweep")
smu.source.level = 7e-3
smu.source.configlist.store("CurrListSweep")
smu.source.level = 11e-3
smu.source.configlist.store("CurrListSweep")
smu.source.level = 9e-3
smu.source.configlist.store("CurrListSweep")
smu.source.sweeplist("CurrListSweep",
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.range = 20
trigger.model.initiate()

Increasing the speed of sweeps

To increase the speed of sweeps:
Reduce the NPLC.
Turn autozero off. If autozero is on, the instrument takes new reference and zero values for every
reading. This can slow down sweep operation. Be aware that if you disable autozero,
measurements may drift and become erroneous. To minimize drift when autozero is disabled, use
the autozero once feature. For more information on the autozero options, see
reference measurements
2450-901-01 Rev. B/September 2013
1, 0.001)
(on page 2-108).
Section 3: Functions and features
Automatic
3-61

Advertisement

Table of Contents
loading

Table of Contents