Section 3: Functions and features
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Recalling a configuration point
Use one of the following methods to recall configuration point 2 on MySourceList:
Using SCPI commands:
:SOURce:CONFiguration:LIST:RECall "MySourceList", 2
Using TSP commands:
smu.source.configlist.recall("MySourceList", 2)
The following figure shows the active state of the instrument after you recall configuration point 2.
Notice that the NPLC setting is 2.0. This is because when you recall a source configuration list, the
settings that are restored are the source settings that were active at the time the point was stored.
Since NPLC is not a source setting, recalling a source configuration list has no effect on the value of
the NPLC setting.
Figure 86: Recall configuration point 2
Viewing configuration list contents
You can ask the instrument to display or print the contents of a specific configuration point by sending
a query. The contents returned include all of the active settings that the instrument saved when you
stored the configuration point.
3-48
2450-901-01 Rev. B/September 2013