Appendix D: Model 2450 in a Model 2400 application
Model 2400 to 2450 differences
You can use a Model 2450 in an existing Model 2400 application. The instruments are mostly
compatible, with the following exceptions:
There is no RS-232 communication port.
The contact check function is not available.
Some commands operate differently or are no longer available. These differences are described
in the following sections.
Model 2400 commands that are supported but operate differently
The command :DISPlay:CNDisplay is supported in the 2400 SCPI command set, but will return
the display to the home screen.
The commands :OUTPut:ENABle[:STATe] and :OUTPut:ENABle:TRIPped? are supported in
the 2400 SCPI command set, but they affect the Model 2450 rear panel INTERLOCK connection
instead of the DIGITAL I/O connection.
D-2
Model 2450 Interactive SourceMeter® Instrument Reference Manual
2450-901-01 Rev. B/September 2013