Advertisement

Section 6: SCPI command reference
:CALCulate[1]:<function>:MATH:MMFactor
This command specifies the scale factor for the y = mx + b math operation.
Type
Command and query
Usage
:CALCulate[1]:<function>:MATH:MMFactor <n>
:CALCulate[1]:<function>:MATH:MMFactor?
<function>
<n>
Details
This command sets the scale factor (m) for an mx + b operation for the selected measurement function.
The mx+b math operation lets you manipulate normal display readings (x) mathematically according to the
following calculation:
y = mx + b
Where:
y is the displayed result
m is a user-defined constant for the scale factor
x is the measurement reading (if you are using a relative offset, this is the measurement with relative
offset applied)
b is the user-defined constant for the offset factor
If you send this command without the <function> parameter, it will set the scale factor for all
functions.
Example
:CALC:VOLT:MATH:FORM MXB
:CALC:VOLT:MATH:MMF 0.80
:CALC:VOLT:MATH:MBF 50
:CALC:VOLT:MATH:STATE ON
Also see
:CALCulate[1]:<function>:MATH:FORMat
:CALCulate[1]:<function>:MATH:MBFactor
:CALCulate[1]:<function>:MATH:STATe
6-10
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The function to which this setting applies:
Current: CURRent[:DC]
Resistance: RESistance
Voltage: VOLTage[:DC]
The scale factor; the valid range is −4294967295 to +4294967295
(on page 6-7)
(on page 6-8)
(on page 6-12)
Where saved
Save settings
Measure configuration list
Set the math function for voltage measurements
to mx+b.
Set the scale factor for voltage measurements to
0.80.
Set the offset factor to 50.
Enable the math function.
2450-901-01 Rev. B/September 2013
Default value
1.000000