smu.source.autorange

This attribute determines if the range is selected manually or automatically for the selected source function or
voltage source.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
sourceAutorange = smu.source.autorange
smu.source.autorange = sourceAutorange
sourceAutorange
Details
This command indicates the state of the range for the selected source. When automatic source range is
disabled, the source range is set manually.
When automatic source range is enabled, the instrument selects the range that is most appropriate for the value
that is being sourced. The output level controls the range. If you read the range after the output level is set, the
instrument returns the range that the instrument chose as appropriate for that source level.
If the source range is set to a specific value from the front panel or a remote command, the setting for automatic
range is set to disabled.
Only available for current and voltage functions.
Example
smu.source.func = smu.FUNC_DC_CURRENT
smu.source.autorange = smu.ON
Also see
smu.source.range
2450-901-01 Rev. B/September 2013
Affected by
Restore configuration
Instrument reset
Power cycle
Source configuration list
Function change
Disable automatic source range: smu.OFF or 0
Enables automatic source range: smu.ON or 1
(on page 8-143)
Section 8: TSP command reference
Where saved
Default value
Configuration script
smu.ON
Source configuration list
Set the source function to current.
Set the instrument to select the
source range automatically.
8-127