Trigger:timer<N>:Delay - Keithley 2450 Reference Manual

Interactive sourcemeter instrument
Table of Contents

Advertisement

Section 6: SCPI command reference
:TRIGger:TIMer<n>:DELay
These commands set and read the timer delay.
Type
Command and query
Usage
TRIGger:TIMer<n>:DELay <interval>
TRIGger:TIMer<n>:DELay?
<n>
<interval>
Details
Each time the timer is triggered, it uses this delay period.
Reading this command returns the delay interval that will be used the next time the timer is triggered.
Example
TRIG:TIM2:DEL 50E-6
Also see
None
6-168
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Trigger timer number (1 to 4)
Delay interval in seconds
Where saved
Save settings
Set trigger timer 2 to delay for 50 µs.
2450-901-01 Rev. B/September 2013
Default value
10e-6 (10 µs)