Section 6: SCPI command reference

:STATus:PRESet

This command resets all bits in the status model.
Type
Command only
Usage
:STATus:PRESet
Details
This function clears the event registers and the enable registers for operation and questionable. It will not clear
the enable status request enable (*SRE) to standard enable (*ESE).
Preset does not affect the event queue.
The Status Event Status Register is not affected by this command.
Example
STAT:PRES
Also see
Status model

:STATus:QUEStionable:CONDition?

This command reads the Questionable Condition Register of the status model.
Type
Query only
Usage
:STATus:QUEStionable:CONDition?
Details
This command reads the contents of the Questionable Condition Register, which is one of the Questionable
Event Registers.
For detail on interpreting the value of a register, see
Example
:STAT:QUES:COND?
Also see
None
6-92
Affected by
Not applicable
(on page C-1)
Affected by
Not applicable
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Where saved
Not applicable
Resets the registers.
Where saved
Not applicable
Understanding bit settings
Reads the Questionable Condition Register.
Default value
Not applicable
Default value
Not applicable
(on page C-15).
2450-901-01 Rev. B/September 2013