Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 3: Functions and features
Setting up a sweep from the front panel
To set up a sweep from the front panel, you select options from the Sweep Settings screen.
Set up the sweep from the front panel
1. Select FUNCTION and select the source and measure functions.
2. On the home screen, set the Source value.
3. Press Menu.
4. Under Source, select Sweep. The Sweep Settings screen is displayed.
Figure 89: Sweep Settings screen - first page
5. Make selections appropriate to your sweep. See the table below for detail on the options.
6. Swipe down to see additional options.
Figure 90: Sweep Settings screen - second page
7. Make selections appropriate to your sweep. See the table below for detail on the options.
8. Select Generate.
9. To run the sweep, press the TRIGGER key.