Smu.source.sweeplog() - Keithley 2450 Reference Manual

Interactive sourcemeter instrument
Table of Contents

Advertisement

Model 2450 Interactive SourceMeter® Instrument Reference Manual

smu.source.sweeplog()

This function creates a logarithmic source sweep configuration list and trigger model.
Type
TSP-Link accessible
Function
Yes
Usage
smu.source.sweeplog(configListName, start, stop, points)
smu.source.sweeplog(configListName, start, stop, points, delay)
smu.source.sweeplog(configListName, start, stop, points, delay, count)
smu.source.sweeplog(configListName, start, stop, points, delay, count, rangeType)
smu.source.sweeplog(configListName, start, stop, points, delay, count, rangeType,
failAbort)
smu.source.sweeplog(configListName, start, stop, points, delay, count, rangeType,
failAbort, dual)
smu.source.sweeplog(configListName, start, stop, points, delay, count, rangeType,
failAbort, dual, bufferName)
smu.source.sweeplog(configListName, start, stop, points, delay, count, rangeType,
failAbort, dual, bufferName, asymptote)
configListName
start
stop
points
delay
count
rangeType
failAbort
dual
bufferName
2450-901-01 Rev. B/September 2013
Affected by
A string that contains the name of the configuration list that the instrument will
create for this sweep
The voltage or current source level at which the sweep starts:
Current: 1 pA to 1.05
Voltage: 1 pV to 210
The voltage or current at which the sweep stops:
Current: 1 pA to 1.05
Voltage: 1 pV to 210
The number of source-measure points between the start and stop values of the
sweep (2 to 1e6); to calculate the number of source-measure points in a sweep, use
the following formula:
Points = [(Stop - Start) / Step] + 1
The delay between measurement points; default is smu.DELAY_AUTO, which
enables autodelay, or a specific delay value from 50 µs to 10,000 seconds, or 0 for
no delay
The number of times to run the sweep; default is 1:
Infinite loop: smu.INFINITE
Finite loop: 1 to 268435455
The source range that is used for the sweep:
Most sensitive source range for each source level in the sweep:
smu.RANGE_AUTO or 0
Best fixed range: smu.RANGE_BEST or 2 (default)
Present source range for the entire sweep: smu.RANGE_FIXED or 1
Abort the sweep if the source limit is exceeded: smu.ON (default)
Complete the sweep if the source limit is exceeded: smu.OFF
Determines if the sweep runs from start to stop and then from stop to start:
Sweep from start to stop only: smu.OFF (default)
Sweep from start to stop, then stop to start: smu.ON
The name of a reading buffer; the default buffers (defbuffer1 or defbuffer2) or
the name of a user-defined buffer; if no buffer is specified, this parameter defaults to
defbuffer1
Section 8: TSP command reference
Where saved
Default value
8-151