Set the displayed digits using the front panel
From the front panel:
1. From the home page, swipe the bottom view until the SETTINGS screen is displayed.
2. Next to Display Digits, select the number.
3. Select the digits to display.
This setting takes effect the next time you make measurements.
Set the displayed digits using SCPI commands
To set number of displayed digits using SCPI commands:
Send the command:
:DISPlay:VOLTage:DIGits <n>
Where <n> is:
3: 3.5 digit resolution
4: 4.5 digit resolution
5: 5.5 digit resolution
6: 6.5 digit resolution
To set the displayed digits for current measurements, replace VOLTage with CURRent. To set it for
resistance measurements, replace VOLTage with RESistance.
Set the displayed digits using TSP commands
To set the number of displayed digits using TSP commands:
Send the commands:
smu.measure.func = mFunction
smu.measure.displaydigits = digits
Where mFunction is:
smu.FUNC_DC_CURRENT: Selects current measurement
smu.FUNC_DC_VOLTAGE: Selects voltage measurement
smu.FUNC_RESISTANCE: Selects ohms measurement
And where digits is the number of digits:
6½ display digits: smu.DIGITS_6_5
5½ display digits: smu.DIGITS_5_5
4½ display digits: smu.DIGITS_4_5
3½ display digits: smu.DIGITS_3_5

Customizing a message for the User Display screen

You can customize the message that is displayed on the User Display screen.
You must use a remote interface to customize the User Display screen.