Scan Configuration (7702 Only) - Keithley SourceMeter 2790 Reference Manual

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

Model 2790 SourceMeter
Memory pattern locations
A total of 40 memory pattern locations are available, M0 through M40. Each location can
store source levels, measurement functions (DCV, Ω2, Ω4 functions only), math function,
units, and channels to close when that memory location is scanned.
Scanning memory pattern locations
Memory pattern locations are scanned in the order that they are listed in the scan list. For
example, assume the following scan list:
(@M1, M4, M3, M5, M2, M6)
The above scan will run in this order: M1 > M4 > M3 > M5 > M2 >M6.

Scan configuration (7702 only)

A scan is configured from the scan configuration menu which is accessed by pressing
SHIFT and then CONFIG.
entering the menu structure you can configure a simple scan, an advanced scan, or reset
the configuration to the default setup for a simple scan. Refer to the flowchart in
Figure 7-3
scan."
klqb
®
Switch System Reference Manual
Figure 7-3
for the following discussions on
Only a sequential scan can be configured from the front panel. For a sequential
scan, the scan proceeds from the lowest numbered channel to the highest.
Non-sequential (backward) scanning is only intended to be performed using
remote programming. Unexpected results may occur if a non-sequential scan is
run from the front panel.
For more information, see
sequential scans," page
shows the basic flowchart to configure a scan. After
"Scan
reset,"
"Scanning fundamentals
7-3.
"Simple
scan," and
"Advanced
Sequential and non-
7-11