G-18

:DATE?

(:CALibration:PROTected:CARD1:DATE?
(:CALibration:PROTected:CARD2:DATE?
Purpose
Format
Response
Description
klqb
Example

:INIT

(:CALibration:PROTected:CARD1:INITiate)
Purpose
Format
Parameters
Description
Example
Model 2790 SourceMeter® Switch System Reference Manual
To request the Model 7751/7752/7753 calibration date.
:cal:prot:card1:date?
:cal:prot:card2:date?
<yyyy>, <mm>, <dd>
The CARD1:DATE? and CARD2:DATE? queries allow you to read back
the calibration date from a Model 7751/7752/7753 in Slot 1 or Slot 2,
respectively.
The card calibration date is automatically set to the Model 2790 real-time clock
date when the card is calibrated.
:CAL:PROT:CARD1:DATE? 1
To initiate Model 7751/7752/7753 calibration procedures.
:cal:prot:card1:init
None
The :INIT command enables Model 7751/7752/7753 calibration
when performing these procedures over the bus. This command must be
sent to the unit after sending the :CODE command, but before
performing Model 7751/7752/7753 calibration.
:CAL:PROT:CARD1:INIT
Request card 1 cal date.
Initiate calibration.