Model 2790 SourceMeter® Switch System Reference Manual
:INIT
(:CALibration:PROTected:INITiate)
Purpose
Format
Parameters
Description
klqb
Example
:LOCK
(:CALibration:PROTected:LOCK)
Purpose
Format
Parameters
Description
klqb
Example
To initiate comprehensive and factory calibration procedures.
:cal:prot:init
None
The :INIT command enables Model 2790 calibration when perform-
ing these procedures over the bus. This command must be sent to the
unit after sending the :CODE command, but before sending any other
calibration command.
The :INIT command should be sent only once before performing either DC,
AC, or factory calibration. Do not send :INIT before each calibration step.
:CAL:PROT:INIT
To lock out comprehensive or manufacturing calibration.
:cal:prot:lock
None
The :LOCK command allows you to lock out both comprehensive and
manufacturing calibration after completing those procedures. Thus,
:LOCK performs the opposite of enabling calibration with the :CODE
command.
To unlock comprehensive calibration, send the :CODE command. To unlock
manufacturing calibration, hold in the OPEN key while turning on the power.
:CAL:PROT:LOCK
Initiate calibration.
Lock out calibration.
G-5