Recommended test equipment

In order to verify Model 7751/7752/7753 accuracy, you will need the equipment summa-
rized in
klqb
t^okfkd
®
Switch System Reference Manual
Table
16-14.
Table 16-14
Recommended equipment for
Model 7751/7752/7753 verification
Keithley Model 2001 Digital Multimeter:
500V: ±47ppm
HP3458A Digital Multimeter:
50mA: ±35ppm
Fluke 5700A Calibrator:
50μA: ±250ppm
Characterized Resistors:
3
Source I ohms:
10Ω
Vishay S102C-10R000-1%
100Ω
Vishay S102C-100R00-1%
50Ω
Vishay S102C-50R000-1%
20Ω
Vishay S102C-20R000-1%
1, 4
Source V ohms:
1GΩ
Caddock MG785-1000M-1%
100MΩ
Caddock MG721-100M-1%
10MΩ
Caddock MG712-10M-1%
1MΩ
Caddock MG712-1M-1%
1. Necessary for Model 7751/7753 only.
2. Characterize all resistors using ohms function of DMM
before use. All resistors are 1% tolerance, 0.5W.
3. Temperature coefficient <20ppm/°C required.
4. Voltage rating of at least 1000V required.
Characterize test resistors using ohms function of either DMM listed in
Table 16-14
before use.
Maximum common-mode voltage (voltage between any terminal and
chassis) is 42V peak.
Hazardous voltages may be present on terminals of test equipment and
test resistors. Use standard safety practices to avoid personal injury
due to electrical shock.