Model 2790 SourceMeter

Data flow (remote operation)

Remote operation can be used with triggering configured to perform a specified number of
measurements and then stop. The various read commands (SENS:DATA?, FETCh?,
READ?, MEAS?, CALC2:DATA?, TRACe:DATA?, and CALC1:DATA?) return the data
array(s) acquired during the measurement cycle. Data flow for this triggering configura-
tion is summarized by the block diagram shown in
for the following discussion.
Figure C-4
Data flow for remote operation
SENSe
Measurement, Filter,
Rel, and Ratio
or Ch Avg
®
Switch System Reference Manual
Sample
Buffer
TRAC:CLE
INIT:CONT OFF
TRIG:COUN 1
INIT
[SENS[1]]:DATA[:LAT]?
[SENS[1]]:DATA:FRESh?
Figure
CALC1
Math (mX+b, m/X+b,
percent, SxIohms, or
SxVohms)
CALC3
Limit Tests
FETCh?
READ?
MEAS?
CALC[1]:DATA[:LAT]?
CALC[1]:DATA:FRESh?
CALC3:LIM1:FAIL?
CALC3:LIM2:FAIL?
C-4. Refer to this block diagram
TRACe
Data Store
CALC2
Min, Max, Sdev,
Mean, Pk-Pk
CALC2:IMM?
CALC2:IMM
CALC2:DATA?
TRACe:DATA?
C-7