Keithley SourceMeter 2790 Reference Manual page 415

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

16-34
3.
4.
5.
6.
7.
8.
Table 16-15
Limits for Model 7751/7752/7753 source-current resistance verification
Source
current
50mA
20mA
10mA
1mA
* Calculate limits based on actual characterized resistance values and Model 7751/7752/7753 resistance
Set the Model 7751/7752/7753 source current to 50mA as follows:
a.
Press SHIFT, CARD, then CONFIG.
b.
Select SLOT1: 7751, SLOT1: 7752, or SLOT1: 7753, then press ENTER.
c.
Select the I-C submenu, then set the current source to 50mA.
d.
Press EXIT to return to normal display.
Using the CLOSE key, close the following channels: 1, 2, 17, 18, and 21.
Set the math function to display resistance as follows:
a.
Press SHIFT then MATH.
b.
Select S1IOHMS, then press ENTER.
c.
Press EXIT to return to normal display.
Recalculate reading limits shown in
resistance value.
Verify that the resistance reading is within the limits calculated in step 6.
Repeat steps 1 through 7 for the 20mA, 10mA, and 1mA source currents, as shown
in
Table
16-15, using the indicated test resistance value for each current. Note:
close channel 24 before the 1mA source test (dry circuit ohms).
Nominal
resistance
20Ω
50Ω
100Ω
10Ω
accuracy specifications. See Verification limits.
Model 2790 SourceMeter
Table 16-15
Nominal reading limits
(1 year, 18°C to 28°C)
19.9814 to 20.0186Ω
49.9435 to 50.0565Ω
99.837 to 100.163Ω
9.86 to 10.14Ω
®
Switch System Reference Manual
based on the actual characterized
Recalculated limits*
__________ to __________ Ω
__________ to __________ Ω
__________ to __________ Ω
__________ to __________ Ω

Advertisement