12-6
E *SAV <NRf> — save
*RCL <NRf> — recall
Parameters
Use the *SAV command to save the present instrument setup configuration in memory for
later recall. Any control affected by *RST can be saved by the *SAV command. The *RCL
command is used to restore the instrument to the saved setup configuration. Four setup
configurations can be saved and recalled. A saved setup is approximately 4k bytes in size.
Model 2790 ships from the factory with SYSTem:PRESet defaults loaded into the avail-
able setup memory. If a recall error occurs, the setup memory defaults to the
SYSTem:PRESet values.
klqb
Programming example:
*SAV 2
*RST
*RCL 2
F *RST — reset
When the *RST command is sent, Model 2790 performs the following operations:
1.
2.
3.
klqb
0 = Memory location 0
1 = Memory location 1
2 = Memory location 2
3 = Memory location 3
For RS-232 operation, *OPC or *OPC? should be used with *SAV and *RCL,
which are slow responding commands. Details on *OPC and *OPC? are pro-
vided in earlier in this section.
' Save present setup in memory location 2.
' Return 2790 to RST defaults.
' Return (recall) 2790 to setup stored in memory location 2.
Returns Model 2790 to the RST default conditions (see "Default" column of SCPI
tables).
Cancels all pending commands.
Cancels response to any previously received *OPC and *OPC? commands.
For RS-232 operation, *OPC or *OPC? should be used with *RST, which is a
slow responding command. Details on *OPC and *OPC? are provided earlier
in this section.
Model 2790 SourceMeter
Save present setup in memory
Return to setup stored in memory
Return Model 2790 to RST defaults
®
Switch System Reference Manual