Model 2790 SourceMeter
Resistance calibration
Follow these steps for resistance calibration:
1.
klqb
2.
Table 17-9
Resistance calibration programming steps
Calibration
1kΩ
10kΩ
100kΩ
1MΩ
*Use exact calibrator resistance value for parameter.
®
Switch System Reference Manual
Set the calibrator to the resistance mode and turn on external sensing.
Use external sense (4-wire Ω) when calibrating all resistance ranges. Be sure
that the calibrator external sense mode is turned on.
Perform the calibration steps summarized in
•
Set the calibrator to the indicated resistance and make sure the unit is in
operate. (Use the recommended resistance or the closest available value.)
•
Send the indicated programming command. (Change the command parameter
if you are using a different calibration resistance than that shown.)
•
Wait until the Model 2790 completes each step before continuing.
Calibrator
step
resistance
1kΩ
10kΩ
100kΩ
1MΩ
Table
Calibration command*
:CAL:PROT:DC:STEP6 1E3
:CAL:PROT:DC:STEP7 10E3
:CAL:PROT:DC:STEP8 100E3
:CAL:PROT:DC:STEP9 1E6
17-9. For each step:
Parameter range
0.9E3 to 1.1E3
9E3 to 11E3
90E3 to 110E3
900E3 to 1.1E6
17-17