Resistance calibration
Follow these steps for resistance calibration:
1.
klqb
2.
Table 17-9
Resistance calibration programming steps
Calibration
1kΩ
10kΩ
100kΩ
1MΩ
*Use exact calibrator resistance value for parameter.
®
Switch System Reference Manual
Set the calibrator to the resistance mode and turn on external sensing.
Use external sense (4-wire Ω) when calibrating all resistance ranges. Be sure
that the calibrator external sense mode is turned on.
Perform the calibration steps summarized in
Set the calibrator to the indicated resistance and make sure the unit is in
operate. (Use the recommended resistance or the closest available value.)
Send the indicated programming command. (Change the command parameter
if you are using a different calibration resistance than that shown.)
Wait until the Model 2790 completes each step before continuing.
Calibrator
step
resistance
1kΩ
10kΩ
100kΩ
1MΩ
Table
Calibration command*
:CAL:PROT:DC:STEP6 1E3
:CAL:PROT:DC:STEP7 10E3
:CAL:PROT:DC:STEP8 100E3
:CAL:PROT:DC:STEP9 1E6
17-9. For each step:
Parameter range
0.9E3 to 1.1E3
9E3 to 11E3
90E3 to 110E3
900E3 to 1.1E6
17-17