External Triggering - Keithley SourceMeter 2790 Reference Manual

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

Model 2790 SourceMeter
Beeper control
The beeper for Hold can be enabled or disabled from the OUTPUT menu as follows:
1.
2.
3.
4.

External triggering

The EX TRIG key selects triggering from three external sources: trigger link, digital I/O,
and the TRIG key. When EX TRIG is pressed, the TRIG annunciator lights and dashes are
displayed to indicate the instrument is waiting for an external trigger. From the front panel,
press the TRIG key to trigger a single reading. Pressing the EX TRIG key again toggles
back to continuous triggers.
The Model 2790 uses two lines of the TRIG LINK rear panel connector as External Trig-
ger (EXT TRIG) input and Voltmeter Complete (VMC) output. The EXT TRIG line
allows the Model 2790 to be triggered by other instruments. The VMC line allows the
Model 2790 to trigger other instruments.
Line 1 is configured as VMC and line 2 as EXT TRIG. The connector pinout is shown in
Figure
Figure 8-3
TRIG LINK pinout
Pin 2
External
Trigger
Input
®
Switch System Reference Manual
Press SHIFT and then OUTPUT.
Use the
or
key to display the present beeper (BEEP) state: NEVER, OUT-
SIDE, or INSIDE.
Perform step a or b:
a.
To enable the beeper, use the
b.
To disable the beeper, use the
Press ENTER. The instrument returns to the normal display state and measurement
state.
8-3.
TRIG LINK
Pinout
8
7
6
5
4
3
2
1
Pin 1
Voltmeter
Complete
Output
or
key to display OUTSIDE or INSIDE.
or
key to display NEVER.
Pin Number
Description
1
Voltmeter Complete Output
2
External Trigger Input
3
No Connection
4
No Connection
5
No Connection
6
No Connection
7
Signal Ground
8
Signal Ground
8-7

Advertisement

Table of Contents
loading

Table of Contents