7751/7753 V-Source - Keithley SourceMeter 2790 Reference Manual

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

Model 2790 SourceMeter Switch System Reference Manual
Ohms temperature coefficient
TC % of reading = (0.001 + 0.025/ISOURCE) % / °C
TC ohms offset = (0.001/ISOURCE) Ω / °C
where: I
Example – Determine temperature coefficient for an I-source setting of 30mA:
% of reading
Ohms offset
Ohms temperature coefficient (30mA) = (0.00183% + 0.033mΩ) / ºC

7751/7753 V-source

Ohms accuracy
High ohms specifications using the V-source of the 7751/7753 are provided in the specifica-
tions for the 500V and 50V settings. Use the graphs in
Figure A-2
is the current output setting in mA
SOURCE
= (0.001 + 0.025/30) % / °C
= (0.001 + 0.00083) %
= 0.00183% / °C
= (0.001/30) Ω / °C
= 0.033mΩ / °C
(7753) for other V-source settings.