Model 2790 SourceMeter

Saving and restoring setups

Saving a user setup
1.
2.
3.
4.
Saving a power-on setup
1.
2.
3.
4.
5.
Restoring a setup
1.
2.
3.
klqb
®
Switch System Reference Manual
Configure Model 2790 for the desired measurement application.
Press SHIFT and then SAVE to access the save setup menu.
Press
to place the cursor on the present setup (SAV0, SAV1, SAV2, SAV3).
Use the
or
key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
Configure Model 2790 for the desired measurement application.
Press SHIFT and then SAVE to access the save setup menu.
Press the
key to display the present power-on (PWR-ON) setup: FACT and
*RST.
Press
to place the cursor on the present power-on setup.
Use the
or
key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
Press SHIFT and then SETUP to access the restore setup menu.
Press
to place the cursor on the present RESTORE setup (FACT or *RST). Only
FACT or *RST is allowed as the power-on setup.
Use the
or
key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
If the settings for a user setup or power-on setup do not match the switching
module types presently installed in the Model 2790, error +520 (Saved setup
scancard mismatch) occurs when the setup is recalled. The scan list will reset to
the factory defaults and all channels will open. However, the saved setup is still
retained in memory and can be restored when the matching switching module is
later installed.