7-16
5.
Advanced scan setup procedure
Step 1: Select the advanced scan configuration menu
1.
2.
Step 2: Edit scan channels
1.
klqb
2.
klqb
Ω 4 function key for a primary channel, the subsequently paired
When you press the
channels will be displayed briefly. Model 7702 example: If you press
108, channels 109 through 120 will also assume the
"128-140 PRD" will be displayed to indicate the paired channels.
A channel that is paired to a primary channel is not affected by function changes.
Model 7702 example: Assume channel 102 is paired to channel 122. Now select channel
103 and press DCV. All following channels, except channel 122, will assume the DCV
setup. Channel 122 remains paired to channel 102. However, if you select channel 101
and press DCV, channel 102 will change to DCV and not be paired to channel 122
anymore. Therefore all 40 channels will assume the DCV setup.
Setting changes — When you press a key to change a setting (i.e., range, rel, digits,
etc.), only the selected scan channel is affected. Model 7702 example: If you make
a range change for channel 103, the range settings for other channels are not
affected.
Press SHIFT and then CONFIG to access the scan setup menu.
Press the
or
key to display INT: ADVANCED and press ENTER.
Use the
or
key to select channel 101:
SETUP
V:101
The CLOSE key can instead be used to select a scan channel to be edited. Press
CLOSE, use the
or
You can disable the channel or use it in the scan. Perform step a or step b:
a.
If you do not want to use the channel, press SHIFT and then CH-OFF to disable
the channel. Available channels that follow will also disable. Note, however, that
channels for slot 2 are not affected.
b.
If you want the channel in the scan, you can either use the presently selected
function or select a valid measurement function. When you press a function
key (i.e., DCV), the channel assumes the setup of the selected function. Avail-
able channels that follow will assume the same setup. Note, however, that
channels for slot 2 are not affected.
For the Model 7702, DCI and ACI cannot be selected for channels 1 through 40.
DCI and ACI are the only functions that can be set for channels 41 and 42.
Model 2790 SourceMeter
(factory default)
key to display the channel, and then press ENTER.
®
Switch System Reference Manual
Ω 4 for channel
Ω 4 function and the message