Remote programming example (multiple channel operation)
The following example assumes a Model 7751 installed in slot 1. This command sequence
connects a DUT at Bank 1 to the DMM input:
ROUT:OPEN:ALL
ROUT:MULT:CLOS (@101,114,118)
The complete simplified schematic of Model 7751/7752/7753 is provided in
When finished with multiple channel operation, it is a good, safe practice to open all chan-
nels (
d.
ROUTe:MULTiple:CLOSe?
This query command returns a <clist> of all closed channels, including
non-measurement channels and paired channels for 4-wire functions.
e.
ROUTe:MULTiple:CLOSe:STATe? <clist>
This query returns a "0" (open) or "1" (closed) for every channel specified in the
<clist>. It is valid for both measurement and non-measurement channels.
For example, assume 7702 channel 245 is closed, and you use this command to
query channels 201, 204, and 245 (<clist> = (@201, 204, 245). The response
message returns "0, 0, 1" to indicate that channels 201 and 204 are open, and
channel 245 is closed.
ROUT:OPEN:ALL
).
Model 2790 SourceMeter
' Open all channels.
' Close channels to connect
channel 1 to DMM.
®
Switch System Reference Manual
Figure
2-12.