Application — sorting resistors
For this application, the idea is to sort a batch of 100Ω resistors into three bins. Bin 1 is for
resistors that are within 1% of the nominal value. Bin 2 is for resistors that exceed 1% tol-
erance, but are within 5%. Bin 3 is for resistors that exceed 5% tolerance.
The digital outputs of the Model 2790 can be used to further automate the test system by
controlling a compatible component handler to perform the binning operations.

Limits

Limit testing is used to test resistor tolerances.
offset-compensated ohms to test 100Ω resistors.
Figure 9-6
Setup to test 100
A) Front panel inputs
Model 2790
B) Model 7702
Ω
resistors
Sense HI
Input HI
Input LO
Sense LO
H
Model 7702
CH 21-40
Switching
L
SENSE
2790 SourceMeter
Figure 9-6
Ω
100
H
CH 1-20
Module
L
INPUT
Ω
100
®
Switch System Reference Manual
shows a basic setup using 4-wire