When the scan is run by pressing STEP or SCAN, channels 1 through 40 will be scanned
and the 40 DCV readings will be stored in the buffer.
Perform the following steps to reset the scan configuration:
1.
2.
After briefly displaying "LIST RESET," the instrument returns to the normal measure-
ment state.

Simple scan

For a simple scan, you specify a starting channel (MIN CHAN) and an ending channel
(MAX CHAN) for the scan. These settings determine the number of channels in the scan.
For example, if you set MIN CHAN to 101 and MAX CHAN to 110, there will be 10
channels in the scan list.
The starting channel number must be lower than the ending channel number. If you enter
an invalid value, the message "TOO SMALL" or "TOO LARGE" will be displayed
briefly. The displayed channel number will default to the lowest available channel (i.e.,
101) or the highest available channel (i.e., 240).
Perform the following steps to configure a simple scan:
1.
2.
3.
4.
5.
6.
7.
8.
Press SHIFT and then CONFIG to enter the scan configuration menu.
Press the
or
key to display INT: RESET and press ENTER.
While in the normal measurement mode, set up the instrument for your test. This
setup will be used for all selected channels in the scan.
Press SHIFT and then CONFIG to access the scan setup menu.
Press the
or
key to display "INT: SIMPLE" and press ENTER.
Set the minimum channel (MIN CHAN) for the scan and press ENTER.
Set the maximum channel (MAX CHAN) for the scan and press ENTER.
Enable (YES) or disable (NO) the timer and press ENTER. See "STEP and SCAN
keys" for details on timer operation.
If you enabled the timer, set the timer interval using the hour/minute/second for-
mat. The timer can be set from 0.001 sec (00H:00M:00.001S) to 99 hrs, 99 min,
99.999 sec (99H:99M:99.999S). Note that pressing the AUTO key sets the timer to
0.001 sec. With the desired interval displayed, press ENTER.
The displayed reading count (RDG CNT) sets the number of channels to scan
(STEP) or the number of scans to run (SCAN). You can change the reading count
to any finite value from 2 to 55000 or you can select infinite (continuous scanning).
To select infinite, set the reading count to 000000 to display INF. See
models" page
7-4, for details on reading count. With the desired reading count set-
ting displayed, press ENTER to return to the normal measurement display state.
Model 2790 SourceMeter
®
Switch System Reference Manual
"Trigger