Switch System Reference Manual
Select the DC volts function by pressing the DCV key and set the Model 2790 to
the 100mV range. Close Channel 1 by pressing the CLOSE key and then keying in
101.
Set the calibrator output to 0.00000mV DC and allow the reading to settle.
Enable the Model 2790 REL mode. Leave REL enabled for the remainder of the
DC volts verification tests.
Source positive, negative, and full-scale voltages for each of the ranges listed in
Table
16-8. For each voltage setting, be sure that the reading is within stated limits.
Press the OPEN key to open Channel 1.
Table 16-8
Model 7702 DCV reading limits
Range
DC voltage*
100mV
100.0000mV
1V
1.000000V
10V
10.00000V
100V
100.0000V
1000V
300.000V
*Source positive and negative values for each range.
Applied
Reading limits
(1 year, 18° to 28°C)
99.990 to 100.010mV
0.999963 to 1.000037V
9.99965 to 10.00035V
99.9936 to 100.0064V
299.976 to 300.024V
16-19