Verification limits

The verification limits stated in this section have been calculated using only the
Model 2790 one-year accuracy specifications, and they do not include test equipment
uncertainty. If a particular measurement falls slightly outside the allowable range, recalcu-
late new limits based on both Model 2790 specifications and pertinent calibration equip-
ment specifications.

Example reading limit calculation

The following is an example of how reading limits have been calculated. Assume you are
testing the 10V DC range using a 10V input value. Using the Model 2790 one-year accu-
racy specification for 10V DC of ± (30ppm of reading + 5ppm of range), the calculated
limits are:
Reading limits = 10V ± [(10V × 30ppm) + (10V × 5ppm)]
Reading limits = 10V ± (0.0003 + 0.00005)
Reading limits = 10V ± 0.00035V
Reading limits = 9.99965V to 10.00035V

Calculating resistance reading limits

Resistance reading limits must be recalculated based on the actual calibration resistance
values supplied by the equipment manufacturer. Calculations are performed in the same
manner as shown in the preceding example, except, of course, that you should use the
actual calibration resistance values instead of the nominal values when performing your
calculations.
For example, assume that you are testing the 10kΩ range using an actual 10.03kΩ calibra-
tion resistance value. Using Model 2790 one-year 10kΩ range accuracy of ± (100ppm of
reading + 6ppm of range), the calculated reading limits are:
Reading limits = 10.03kΩ ± [(10.03kΩ × 100ppm) + (10kΩ × 6ppm)]
Reading limits = 10.02894kΩ to 10..03106kΩ

Restoring factory defaults

Before performing the verification procedures, restore the instrument to its factory
defaults as follows:
1.
2.
®
Switch System Reference Manual
Press SHIFT and then SETUP. The instrument will display the following prompt:
RESTORE: FACT.
Using either range key, select FACT, then restore the factory default conditions by
pressing ENTER.
16-5