G-24

Error queue

As with other Model 2790 errors, any calibration error will be reported in the bus error
queue. You can read this queue by using the :SYST:ERR? query. The Model 2790 will
respond with the appropriate error number and message, as summarized in Table G-5.

Status byte EAV (Error Available) bit

Whenever an error is available in the error queue, the EAV (Error Available) bit (bit 2) of
the status byte will be set. Use the *STB? query or serial polling to obtain the status byte,
then test bit 2 to see if it is set. If the EAV bit is set, an error has occurred and you can use
the :SYST:ERR? query to read the error and at the same time clear the EAV bit in the sta-
tus byte. Bit 3 of the Standard Event Register will also be set by a calibration error.

Generating an SRQ on error

To program the instrument to generate an IEEE-488 bus SRQ when an error occurs, send
the following command: *SRE 4. This command will enable SRQ when the EAV bit is set.
You can then read the status byte and error queue as outlined above to check for errors and
to determine the exact nature of the error.

Detecting calibration step completion

When sending remote calibration commands, you must wait until the instrument com-
pletes the current operation before sending a command. You can use either *OPC? or
*OPC to help determine when each calibration step is completed.

Using the *OPC? query

With the *OPC? (operation complete) query, the instrument will place an ASCII 1 in the
output queue when it has completed each step. To determine when the OPC response is
ready, do the following:
1.
2.
3.
Repeatedly test the MAV (Message Available) bit (bit 4) in the status byte and wait
until it is set. (You can request the status byte by using the *STB? query or by serial
polling.)
When MAV is set, a message is available in the output queue and you can read the
output queue and test for an ASCII 1.
After reading the output queue, repeatedly test MAV again until it clears. At this
point, the calibration step is completed.
Model 2790 SourceMeter® Switch System Reference Manual