Reading Relay Closure Count; Resetting Relay Closure Count - Keithley SourceMeter 2790 Reference Manual

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

18-10

Reading relay closure count

To determine the closure count of specific channels, send this query via remote:
:ROUTe:CLOSe:COUNt? (@clist)
Here, clist is the summary of channels. For example, to determine the closure count of
channels 1 and 4 of a module in slot 1, the following query would be sent:
:ROUT:CLOS:COUN? (@101,104)
The following query would determine the closure count of slot 1 module channels 1
through 10:
:ROUT:CLOS:COUN? (@101:110)

Resetting relay closure count

klqb
To reset the relay closure count of specific channels to 0, send this command via remote:
:ROUTe:CLOSe:RCOunt (@clist)
Again clist is the summary of channels to be reset. For example, the following command
resets channels 2 and 7 of a module in slot 1 to 0:
:ROUT:CLOS:RCO (@102,107)
The following command resets the count of slot 1 module channels 1 through 10:
:ROUT:CLOS:RCO (@101:110)
Model 2790 SourceMeter
The Model 2790 must be in the manufacturing calibration mode to reset the
closure count. To do so, press and hold the OPEN key while turning on the
power, then send the :CAL:PROT:CODE "code" to unlock calibration (default
code: KI002790). After resetting relay counts, send :CAL:PROT:LOCK to lock
out calibration.
®