Remote Calibration - Keithley SourceMeter 2790 Reference Manual

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

17-14

Remote calibration

Follow the steps in this section to perform comprehensive procedures via remote. See
Appendix G
When sending calibration commands, be sure that the Model 2790 completes each step
before sending the next command. You can do so either by observing the front panel
CALIBRATING message or by detecting the completion of each step over the bus.
(See
The procedures for calibrating the Model 2790 via remote include:
klqb
Preparing the Model 2790 for calibration
1.
2.
3.
for a detailed list and description of remote calibration commands.
"Detecting calibration step
Preparing the Model 2790 for calibration
Front panel short and open calibration
DC volts calibration
Resistance calibration
DC current calibration
AC volts calibration
AC current calibration
Programming calibration dates
Saving calibration constants
Locking out calibration
As with front panel calibration, you can choose to perform comprehensive,
DC-only, or AC-only calibration. Be sure to include a space character between
each command and parameter.
Connect the Model 2790 to the IEEE-488 bus of the computer using a shielded
IEEE-488 cable, such as the Keithley Model 7007, or connect the unit to a com-
puter through the RS-232 port using a straight-through 9-pin to 9-pin cable (use a
9-25-pin adapter if necessary).
Turn on the Model 2790 and allow it to warm up for at least two hours before per-
forming calibration.
Make sure the primary address of the Model 2790 is the same as the address speci-
fied in the program that you will be using to send commands. (Use the GPIB key.)
Model 2790 SourceMeter
completion" in
Appendix
®
Switch System Reference Manual
G.)

Advertisement

Table of Contents
loading

Table of Contents