Verifying Ac Current - Keithley SourceMeter 2790 Reference Manual

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

Model 2790 SourceMeter

Verifying AC current

Check AC current accuracy by applying accurate AC voltage current at specific frequencies
from the AC current calibrator to the Model 2790 input and verifying that the displayed
readings fall within specified limits. Follow these steps to verify AC current:
1.
Figure 16-4
Connections for Model 2790 AC current verification
Model 2790
Model 2790 SourceMeter Switch System
MATH O U T P U T
RATIO
CH AVG
CONT
OCOMP
PERIOD SENSOR
Ω2
Ω4
SHIFT
DCV
ACV
DCI
ACI
FREQ
DELAY
HOLD
LIMITS
ON/OFF
TYPE
MONITOR
CH-OFF
LOCAL
EX TRIG
TRIG
STORE
RECALL
FILTER
REL
POWER
SAVE
SETUP
CONFIG
HALT
TEST
LSYNC
GPIB
OPEN
CLOSE
STEP
SCAN
DIGITS RATE
EXIT
2.
3.
4.
®
Switch System Reference Manual
Connect the Model 2790 AMPS and INPUT LO jacks to the calibrator as shown in
Figure
16-4. Be sure the INPUTS switch is in the FRONT position.
INPUT LO
SENSE
INPUT
Ω 4 WIRE
HI
350V
1000V
!
PEAK
PEAK
LO
500V
PEAK
INPUTS
TEMP
RANGE
F
R
F
F
CARD
AUTO
FRONT/REAR
3A 250V
AMPS
RANGE
RS-232
ENTER
AMPS
Select the AC current function by pressing the ACI key.
Set the Model 2790 for the 1A range.
Source 1A and 3A, 1kHz full-scale AC currents as summarized in
verify that the readings are within stated limits.
Table 16-5
ACI limits
Applied
ACI range
AC current
1A
1.000000A
3A
3.00000A*
*If the Fluke 5725A amplifier is not available, apply 2.2A from the
calibrator. Reading limits for 2.2A are 2.1949 to 2.2051A.
OUTPUT
HI
OUTPUT LO
Calibrator (Output AC Current)
Reading limits @ 1kHz
(1 year, 18°C to 28°C)
0.99860 to 1.00140A
2.9937 to 3.0063A
16-13
Table 16-5