Model 2790 SourceMeter
Scanning
The moving filter cannot be used when scanning. A scan channel cannot be configured to
use the moving filter. Also, the filter window is not used when scanning.
When a simple scan is configured, the present filter count and state will apply to all chan-
nels in the scan. The window setting is ignored (effectively set to NONE) and, if the mov-
ing filter is selected, the filter will not enable when the scan is run.
For the advanced scan, filter state (on or off) and count can be set for each channel. You
cannot set unique filter count, type, and window settings from the advanced scan setup
menu.
For remote programming, the <clist> parameter is used to set filter count and state for
each 7702 channel in the scan. You cannot set unique filter type and window settings.