5-14

Basic operation

klqb
1.
2.
3.
4.
Front panel scanning (7702 only)
When a simple scan using a 7702 module is configured, the present math calculation will
apply to all channels in the scan. When an advanced scan is configured, each channel can
have its own unique math setup. Details to configure and run a scan are provided in
Section
Remote scanning
For remote programming, the <clist> parameter is used to configure channels for a 7702
scan
allow scanning through 7751/7752/7753 channels (see
If using switching module inputs, make sure the front panel INPUTS switch is set
to the REAR position (in). If using the front panel inputs, the switch must be in
the FRONT position (out).
Select and (if required) configure the desired math function as previously
explained.
Select the desired measurement function.
Apply the signal to be measured to a switching channel input or to the front panel
inputs.
If using a switching module (REAR inputs selected), close the appropriate chan-
nel(s) to connect the input to the DMM. If using the front panel inputs (FRONT
inputs selected), it does not matter if a switching channel is closed. The result of
the math calculation will be displayed.
7.
(Section
7). The <clist> parameter can also include memory pattern locations to
Model 2790 SourceMeter
®
Switch System Reference Manual
"Memory patterns," page
2-36).