Keithley SourceMeter 2790 Reference Manual page 426

Hide thumbs Also See for SourceMeter 2790:
Table of Contents

Advertisement

Model 2790 SourceMeter
3.
Table 17-2
Comprehensive calibration procedures
Full calibration
DCV, DCI, and ohms
ACV and ACI
Front panel short and open calibration
At the Model 2790 prompt for a front panel short, do the following:
1.
Figure 17-1
Low-thermal short connections
SHIFT
LOCAL
POWER
klqb
2.
®
Switch System Reference Manual
Choose which of the calibration tests summarized in
the CAL: RUN prompt. Use the up and down range keys to scroll through the
options; select your choice by pressing ENTER.
Menu
Procedure
choice
ALL
DC
AC
Connect the Model 8610 low-thermal short to the instrument front panel INPUT
and SENSE terminals as shown in
not pressed in so that the front inputs are selected. Wait at least three minutes
before proceeding to allow for thermal equilibrium.
Model 2790
Model 2790 SourceMeter Switch System
RATIO
CH AVG
CONT
OCOMP
MATH O U T P U T
Ω2
Ω4
DCV
ACV
DCI
ACI
MONITOR
DELAY
HOLD
LIMITS
ON/OFF
TYPE
EX TRIG
TRIG
STORE
RECALL
FILTER
REL
SAVE
SETUP
CONFIG
HALT
TEST
LSYNC
CLOSE
OPEN
STEP
SCAN
DIGITS RATE
Be sure to connect the low-thermal short properly to the HI, LO, and SENSE
terminals. Keep drafts away from low-thermal connections to avoid thermal
drift, which could affect calibration accuracy.
Press ENTER to start short-circuit calibration. While the unit is calibrating, it will
display:
CALIBRATING
All comprehensive calibration steps (DC and AC).
DC voltage, DC current, and resistance calibration.
AC voltage and AC current calibration.
Figure
17-1. Make sure the INPUTS button is
S+
HI
SENSE
INPUT
Ω 4 WIRE
HI
350V
1000V
!
PEAK
PEAK
LO
LO
500V
PERIOD SENSOR
PEAK
INPUTS
FREQ
TEMP
RANGE
F
F
R
F
CH-OFF
CARD
AUTO
FRONT/REAR
3A 250V
AMPS
RANGE
GPIB
RS-232
EXIT
ENTER
S-
Table 17-2
you want to run at
Procedures
Model 8610
Low-thermal
short
LO
17-7